SH-IVD

- Wide range (0 – 10ppm to 0 – 100%, Selection of 2 ranges out of 18 ranges)
- 2 patterns of portable and process
- LED digital display with versatile display functions
- Equipped with self-diagnosis function
- Simple, one-push calibration
Use
- Air separation plants
- Ferrite incinerators
- Oxygen enriching equipment
- Gas displacement monitoring
- Gas purity control (H2, He, Ar, ...)
- Heat treatment of iron and steel and other metals (non-oxidation furnaces)
- Special metal welding atmosphere gas (welding process)
- Semiconductor process atmosphere gas
- Inactive gas for seals
Specifications
| Model | SH-IVD |
|---|---|
| External dimensions | 202(W)×165(H)×342(D) |
| Panel cut | 190(W)×150(H) |
| Painting color | Case: Munsell N3 semigloss, Panel: Munsell 5Y7/1 semigloss |
| Range | Selection of 2 ranges out of 16 ranges (0 – 1, 2, 5, 10, 25, 100%O2) (0-10, 20, 50, 100, 200, 500, 1,000, 2,000, 5,000ppm) |
| Output | 0 – 10mV or 0 – 1V, 4 – 20mA (Isolate linear output) |
| Sample gas flow rate | 500ml/min |
| Sample gas pressure | 3 – 100kPaG (Sample gas flow rate at 500ml/min) Pump required at ≦3kPaG, Pressure reducing valve required at ≧100 kPaG |
| Sample gas conditions | No inclusion of combustible gas or corrosive gas, No condensation at ambient temperature |
| Reproducibility precision | % range: ±0.5%FS, ppm range: ±1%FS |
| Response time | % range: Within 5 sec (90% response), ppm range: Within 10 sec (90% response) |
| Ambient temperature | -5 to 40°C |
| Power voltage | 100VAC±10%, 50/60Hz, 200VA |
| Weight | Approx. 7kg |
TF-SD

Sensor Model AS-210

Receiver Model DTF-201
- Separation enabled between the sensor and the receiver
- Stand-alone panel type available with the combination of sampling materials
- No need of standards gas or reference air (air 1-point calibration)
- Reduction zone measurement type also available
Use
Measurement of oxygen concentration in oxidized combustion exhaust gas
- Small- and medium-sized boilers
- Small- and medium-sized industrial heating furnaces
- Coke ovens
- Heat-treating furnaces (radiant tube furnaces)
Measurement of atmosphere concentration in the reduced combustion exhaust gas
- Ceramic kilns (shuttle furnaces)
- Ceramic kilns (tunnel furnaces)
- Ceramic raw material firing furnaces (rotary kilns)
Specifications
| Model | TF-SD | |
|---|---|---|
| Measurement principle | Zirconia 2-cell form | |
| Measurement range | 0 – 5/10/25% , -15/0/25%, -6/10%, -3/5% | |
| Output | 1st output | Selection from among 4 – 20mADC |
| 2nd output | Selection from among 0-1 VDC / 0-5 VDC / 0-10 VDC | |
| Response time | Within 10 sec (90% response at calibration gas changeover) | |
| Alarm display | Content display on the receiver with Error No. | |
| Warming-up time | 3 min | |
| Power source | 100V (85 – 132V) AC, 50/60Hz, Approx. 50VA | |
| External dimensions | 96(W)×96(H)×170(D)(Receiver) | |
TF-21/SH-30 Series

Sensor Model TF-21 / Receiver Model DTF-201

Model SH-304
- Separation enabled between the sensor and the receiver
- Light-weight and compact enough to be build in the system
- Low concentration (applicable to 0 – 10ppm range)
Use
- Semiconductor manufacturing equipment: Diffusion furnaces, RTP, measurement of various atmospheres, air separators
Specifications
| Model | TF-21D201 |
|---|---|
| Sensor | TF-21 |
| Receiver | DTF-201 |
| Measurement principle | Zirconia concentration cell form |
| Measurement range | Min. 0 – 10ppm to Max. 0 – 100%, Zero-point fixed free 4-range |
| Measurement coverage | 0 – 100% |
| Calibration | Air point + Max. 4-point calibration enabled |
| Output | 4 – 20mA×1-point (Non-insulated output, Load resistance: ≦600Ω) 0 – 1V / 0 – 5V / 0 – 10V×1-point selection (Non-insulated output) |
| Flow rate of sample gas and calibration gas | 0.5±0.02L/min. |
| Reproducibility | ±1%F.S. |
| Linearity | 0 – ≧100ppm: ±2%FS, 0 – <100ppm: ±3% FS |
| Response time | 0 – ≧1%: ≦10 sec, 0 – <1%: ≦30 sec *90% response at calibration gas changeover |
| Drift | ≦±2%FS/week |
| Power supply | 85 – 132VAC *Option: 170 – 264VAC, 24V DC±10% |
| Contact input (no-voltage contact input) |
Range changeover input |
| Contact output (Open collector output, 25VDC, Max. 50mA, NO/CN changeover enabled) |
Analyzer error Range echo back (2-range identification) Range echo back (3-range identification) Ready Concentration UL alarm (0.0 – 9990.0ppm or 0.0 – 100.0% setting enabled) Concentration LL alarm (0.0 – 9990.0ppm or 0.0 – 100.0% setting enabled) *Selection of 4 points from above, Setting change enabled |
| Range changeover | Local range changeover / Remote range changeover / Auto range changeover *Selection of 4 points from above, Setting change enabled |
| First-order lag setting | 0 – 99 sec setting enabled |
| Output hold setting | None or any value or hold with the value 5 sec before error occurrence *In warming-up, hold in case of the sensor or receiver error occurrence |
| Concentration UL/LL alarm | UL alarm: 0.0 – 9990.0ppm or 0.0 – 100.0% setting enabled LL alarm: 0.0 – 9990.0ppm or 0.0 – 100.0% setting enabled |
| RS232C communication function | Transmitting measurement status, error code, measurement range and measured value Communication cable specifications: Straight cable |
| Model | SH-304 |
|---|---|
| Sensor | TF-21 |
| Receiver | DTF-201 |
| Measurement principle | Zirconia concentration cell form |
| Measurement range | 0 – 10 / 100 / 1,000ppm, 0 – 100% *Min. 0 – 10ppm to Max. 0 – 100% free 4-range |
| Measurement coverage | 0 – 100% |
| Calibration | Air point + Max. 4-point calibration enabled |
| Output | 4 – 20mA×1-point (Non-insulated output, Load resistance: ≦600Ω) 0 – 1V×1-point selection (Non-insulated output) |
| Flow rate of sample gas and calibration gas | 0.5±0.02L/min. |
| Reproducibility | ±1%F.S. |
| Linearity | 0 – ≧100ppm: ±2%FS, 0 – <100ppm: ±3% FS |
| Response time | 0 – ≧1%: ≦10 sec, 0 – <1%: ≦30 sec *90% response at calibration gas changeover |
| Drift | ≦±2%FS/week |
| Power supply | 100VAC *Option: 170 – 264VAC, 24V DC±10% |
| Contact input (no-voltage contact input) |
Range changeover input |
| Contact output (Open collector output, 25VDC, Max. 50mA, NO/CN changeover enabled) |
Analyzer error Ready Concentration UL alarm (0.0 – 9990.0ppm or 0.0 – 100.0% setting enabled) Concentration LL alarm (0.0 – 9990.0ppm or 0.0 – 100.0% setting enabled) Range echo back (2-range identification) Range echo back (3-range identification) *Selection of 4 points from above on standard, Setting change enabled |
| Range changeover | Auto range changeover *Remote range / local range changeover selection, Setting change enabled |
| First-order lag setting | 0 – 99 sec setting enabled |
| Output hold setting | Hold with the value of 5 sec before error occurrence *In warming-up, hold in case of the sensor or receiver error occurrence *No hold or hold to any value is selectable. |
| Concentration UL/LL alarm | UL alarm: 0.0 – 9990.0ppm or 0.0 – 100.0% setting enabled LL alarm: 0.0 – 9990.0ppm or 0.0 – 100.0% setting enabled |
| RS232C communication function | Transmitting measurement status, error code, measurement range and measured value Communication cable specifications: Straight cable |
TF-30/SH-20 Series

Sensor Model TF-30 / Receiver Model DTF-201

Model SH-204
- Separation enabled between the sensor and the receiver
- Light-weight and compact enough to be build in the system
- No need of standards gas or reference air (air 1-point calibration)
Use
- Reflow furnaces
- Ferrite firing furnaces
- Control of other atmospheres and combustion tubes
Specifications
| Model | TF-30D201 |
|---|---|
| Sensor | TF-30 |
| Receiver | DTF-201 |
| Measurement principle | Zirconia 2-cell pump limit current form |
| Measurement range | Min. 0 – 100ppm to Max. 0 – 100%, Zero-point fixed free 4-range *Measurement of ≧25% O2 is disabled. |
| Measurement coverage | 0 – 25% |
| Calibration | Calibration gas air point + Max. 3-point calibration Air 1-point calibration (0 – ≧1,000ppm) Calibration by data input (0 – ≧1,000ppm at sensor replacement) |
| Output | 4 – 20mA×1-point (Non-insulated output, Load resistance: ≦600Ω) 0 – 1V / 0 – 5V / 0 – 10V×1-point selection (Non-insulated output) |
| Flow rate of sample gas and calibration gas | 0.5±0.1L/min. |
| Reproducibility | ±1%FS For ±0 – 25% fixed range: ±0.5FS |
| Linearity | ±2%FS For ±0 – 25% fixed range:±1.0% FS |
| Response time | 0 – ≧1,000ppm: ≦30 sec, 0 – <1,000ppm: ≦60 sec *90% response at calibration gas changeover |
| Drift | ≦±2%FS/week |
| Supply power | 85 – 132VAC *Option: 170 – 264VAC, 24V DC±10% |
| Contact input (no-voltage contact input) |
Range changeover input |
| Contact output (Open collector output, 25VDC, Max. 50mA, NO/CN changeover enabled) |
Analyzer error Range echo back (2-range identification) Range echo back (3-range identification) Ready Concentration UL alarm (0.0 – 9990.0ppm or 0.0 – 100.0% setting enabled) Concentration LL alarm (0.0 – 9990.0ppm or 0.0 – 100.0% setting enabled) *Selection of 4 points from above, Setting change enabled |
| Range changeover | Local range changeover / Remote range changeover / Auto range changeover *Selection from above, Setting change enabled |
| First-order lag setting | 0 – 99 sec setting enabled |
| Output hold setting | None or any value or hold with the value 5 sec before error occurrence *In warming-up, hold in case of the sensor or receiver error occurrence |
| Concentration UL/LL alarm | UL alarm: 0.0 – 9990.0ppm or 0.0 – 100.0% setting enabled LL alarm: 0.0 – 9990.0ppm or 0.0 – 100.0% setting enabled |
| RS232C communication function | Transmitting measurement status, error code, measurement range and measured value Communication cable specifications: Straight cable |
| Type | SH-204 |
|---|---|
| Sensor | TF-30 |
| Receiver | DTF-201 |
| Measurement principle | Zirconia 2-cell pump limit current form |
| Measurement range | Min. 0 – 100ppm to Max. 0 – 100%, Zero-point fixed free 4-range *Measurement of 25% O2 is disabled. |
| Measurement coverage | 0 – 25% |
| Calibration | Air point + Max. 3-point calibration Air 1-point calibration (0 – ≧1,000ppm) Calibration by data input (0 – ≧1,000ppm at sensor replacement) |
| Output | 4 – 20mA×1-point (Non-insulated output, Load resistance: ≦600Ω) 0 – 1V / 0 – 5V / 0 – 10V×1-point selection (Non-insulated output) |
| Flow rate of sample gas and calibration gas | 0.5±0.1L/min. |
| Reproducibility | ±1%FS For ±0 – 25% fixed range: ±0.5FS |
| Linearity | ±5%FS For ±0 – 25% fixed range: ±1.0FS |
| Response time | 0 – ≧1,000ppm: ≦30 sec, 0 – <1,000ppm: ≦60 sec *90% response at calibration gas changeover |
| Drift | ≦±2%FS/week |
| Power Supply | 85 – 132VAC *Option: 170 – 264VAC, 24V DC±10% |
| Contact input (no-voltage contact input) |
Range changeover input |
| Contact output (Open collector output, 25VDC, Max. 50mA, NO/CN changeover enabled) |
Analyzer error Range echo back (2-range identification) Range echo back (3-range identification) Ready Concentration UL alarm (0.0 – 9990.0ppm or 0.0 – 100.0% setting enabled) Concentration LL alarm (0.0 – 9990.0ppm or 0.0 – 100.0% setting enabled) *Selection of 4 points from above, Setting change enabled |
| Range changeover | Local range changeover / Remote range changeover / Auto range changeover *Selection from above, Setting change enabled |
| First-order lag setting | 0 – 99 sec setting enabled |
| Output hold setting | None or any value or hold with the value 5 sec before error occurrence *In warming-up, hold in case of the sensor or receiver error occurrence |
| Concentration UL/LL alarm | UL alarm: 0.0 – 9990.0ppm or 0.0 – 100.0% setting enabled LL alarm: 0.0 – 9990.0ppm or 0.0 – 100.0% setting enabled |